代码如下:

#include "C:\Program Files\IVI Foundation\VISA\WinNT\agvisa\include\visa.h"

#pragma comment(lib,"C:\\Program Files\\IVI Foundation\\VISA\\WinNT\\agvisa\\lib\\msc\\agvisa32.lib")

Char buf[255];

ViSession vi;

ViSession vi_FSP;

//打开频谱仪FSP7

viOpenDefaultRM(&vi);

if(VI_SUCCESS != viOpen(vi,GPIB0::28::INSTR, VI_NULL,VI_NULL, &vi_FSP)){

    //仪器打开失败

Return;

}

viPrintf(vi_FSP,"*CLS\n");

viPrintf(vi_FSP,"SYST:DISP:UPDate ON\n");

viPrintf(vi_FSP,"SYST:PSAVE ON\n");

viPrintf(vi_FSP,"DISP:WIND:TRAC:Y:RLEV -10DBM\n");

viPrintf(vi_FSP,"DISP:TRAC:Y:RLEV:OFFS 0\n");

viPrintf(vi_FSP,"FREQ:SPAN 20MHZ\n");

viPrintf(vi_FSP,"FREQ:CENT 999MHZ\n");

viPrintf(vi_FSP, "BAND:VID:AUTO ON\n");//BAND:VID 300kHz

viPrintf(vi_FSP, "BAND:RES:AUTO ON\n");

viPrintf(vi_FSP,"SWE:TIME 1s\n");

viPrintf(vi_FSP,"INPUT:ATT 30dB\n");

viPrintf(vi_FSP,"BAND:RES 500KHz\n");//500

viPrintf(vi_FSP,"BAND:VID 500kHz\n");//10MHz

viPrintf(vi_FSP,"INIT;*WAI\n");

//----------------------------------

viPrintf(vi_FSP,"FREQ:CENT 999MHZ\n");

viPrintf(vi_FSP,"INIT:CONT OFF\n");

viPrintf(vi_FSP,"INIT:IMM;*WAI\n");

viPrintf(vi_FSP,"CALC:MARK:MAX\n");viClose(vi);

viQueryf(vi_FSP,"CALC:MARK:Y?\n","%t",buf);

viPrintf(vi_FSP,"INIT:CONT ON\n");

//-----------------------------------

Float f_max = atof(buf);

viClose(vi_FSP);

/////////////////////////////////////////////////////

Continuous vs. Single Sweep 

In manual operation, the instrument typically performs measurements repeatedly, i.e. it operates in continuous sweep mode. This is useful for alignment tasks, where the instrument display immediately reflects adjustments performed on the device under test (DUT). 

For remote control operations, however, continuous sweep presents several issues: 